Confocal Fiber Displacement Sensor

ZW-8000 / 7000 / 5000 Series

Reliable measurements for any material and surface types

Sensor Head

ZW-S8010/S8020/S8030/SP8007/SP8010/SPR8007/SPR8010

Item

Specifications

ZW-S8010

ZW-S8020

ZW-S8030

Sensor controller

ZW-8000T

Sensor head type

Square-shaped straight type

Measurement center distance *1

10 mm

20 mm

30 mm

Measuring range *2

±0.5 mm

±1mm

±2mm

Static resolution *3

0.25 μm

Linearity *4

±0.3 μm

±0.6 μm

±1.3 μm

Spot diameter (Total measurent
 range) *5

4 μm dia.

7 μm dia.

10 μm dia.

Measurement cycle *6

60 μs to 7,500 μs

Operating ambient illumination

Illumination on object surface max.30000 Lx: (incandescent light)

Ambient temperature range

Operation: 0 to 50°C, Storage: -15 to +60°C
 (No freezing and condensation)

Ambient humidity range

Operation/storage: 35 or 85%RH (No condensation)

Degree of protection

IP40 (IEC60529)

Vibration resistance (destructive)

10 to 150 Hz (half amplitude 0.35 mm), 80 mins in each of X/Y/Z directions

Shock resistance (destructive)

150 m/s2, 6 direction, 3 times each (up/down, left/right, forward/backward)

Temperature characteristic *7

0.6 μm/°C
 (0.2 μm/°C)

1.1 μm/°C
 (0.5 μm/°C)

1.8 μm/°C
 (1.0 μm/°C)

LED Safety

Risk Group 1 (IEC62471)

LASER safety

Class1 (IEC/EN60825-1)

Material

Chassis: aluminum die cast
 Fiber cable sheath: PVC
 Calibration ROM: PC

Fiber cable length

0.3 m, 2 m (flex-resistant cable)

Fiber cable minimum bend radius

20 mm

Insulation resistance
 (Calibration ROM)

Between case and all terminals: 20 MΩ (by 250 VDC)

Dielectric strength
 (Calibration ROM)

Between case and all terminals: 1000 VAC, 50/60 Hz, 1 min

Weight

Fiber cable length 0.3 m Approx. 170g
 Fiber cable length 2 m Approx. 180g

Accessories

Calibration ROM fixing screw (M2×5mm) × 1,
 Fiber cable protective cap × 1, Strap × 1,
 Instruction Manual, Precautions

Item

Specifications

ZW-SP8007

ZW-SP8010

ZW-SPR8007

ZW-SPR8010

Sensor controller

ZW-8000T

Sensor head type

Pen-shaped straight type

Pen-shaped right angle type

Measurement center distance *1

7 mm

10 mm

7 mm

10 mm

Measuring range *2

±0.3 mm

±0.7 mm

±0.3 mm

±0.7 mm

Static resolution *3

0.25 μm

Linearity *4

±0.3 μm

±0.45 μm

±0.45 μm

±0.7 μm

Spot diameter (Total measurent
 range) *5

7 μm dia.

10 μm dia.

8 μm dia.

11 μm dia.

Measurement cycle *6

60 μs to 7,500 μs

Operating ambient illumination

Illumination on object surface max.30000 Lx: (incandescent light)

Ambient temperature range

Operation: 0 to 50°C, Storage: -15 to +60°C
 (No freezing and condensation)

Ambient humidity range

Operation/storage: 35 or 85%RH (No condensation)

Degree of protection

IP40 (IEC60529)

Vibration resistance (destructive)

10 to 150 Hz (half amplitude 0.35 mm), 80 mins in each of X/Y/Z directions

Shock resistance (destructive)

150 m/s2, 6 direction, 3 times each (up/down, left/right, forward/backward)

Temperature characteristic *7

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

LED Safety

Risk Group 1 (IEC62471)

LASER safety

Class1 (IEC/EN60825-1)

Material

Chassis: SUS
 Fiber cable sheath: PVC
 Calibration ROM: PC
 Mounting Plate: Aluminum

Chassis: SUS, aluminum
 Fiber cable sheath: PVC
 Calibration ROM: PC
 Mounting Plate: Aluminum

Fiber cable length

0.3 m, 2 m (flex-resistant cable)

Fiber cable minimum bend radius

20 mm

Insulation resistance
 (Calibration ROM)

Between case and all terminals: 20 MΩ (by 250 VDC)

Dielectric strength
 (Calibration ROM)

Between case and all terminals: 1000 VAC, 50/60 Hz, 1 min

Weight

Fiber cable length 0.3 m Approx. 27 g
 Fiber cable length 2 m Approx. 37 g

Fiber cable length 0.3 m Approx. 31 g
 Fiber cable length 2 m Approx. 41 g

Accessories

Installation plate × 1, Unit fixing screws (M2 × 10 mm) × 4,
 Calibration ROM fixing screw (M2 × 5 mm) × 1,
 Fiber cable protective cap × 1, Strap × 1,
 Instruction Manual, Precautions







1. Indicates the distance from the front of the sensor head. The pen-shaped right angle type has a maximum individual
       difference of ±0.15 mm in the distance from the front of the sensor head.
 2. The measurement range is higher 100 μs than measurement cycle.
 3. Capacity value when OMRON standard mirror surface target is measured at the measurement center distance as the
       average of 16,384 times.
       The value when the Sensor Controller ZW-8000T is connected.
 4. Material setting for the OMRON standard mirror surface target: Error from an ideal straight line when measuring on
       mirror surface.
5. Capacity value defined by 1/e2 (13.5%) of the peak optical intensity of the measurement wavelength.
 6. When an extension fiber cable of 2 m or longer is connected, the setting rage of the measurement cycle (exposure
       time) changes. For details, refer to Setting Measurement Cycle in the ZW-8000/7000/5000 User's Manual (Cat. No.
       Z362).
 7. Actual value of the change in measurement value at the measurement center distance when fastened with an
      aluminum jig between the Sensor Head and the target, and with the Sensor Head and the Sensor Controller set in the
      same temperature environment.
      The value in parentheses is the actual value when using an SUS304 jig.
      When measuring the thickness, the value is calculated from the difference between the heights of the surface and rear
      surface, so there is no effect on the temperature change.

ZW-S7010/S7020/S7030/S7040/SP7007/SP7010/SPR7007/SPR7010

Item

Specifications

ZW-S7010

ZW-S7020

ZW-S7030

ZW-S7040

Sensor controller

ZW-7000T

Sensor head type

Square-shaped straight type

Measurement center distance *1

10 mm

20 mm

30 mm

40 mm

Measuring range *2

±0.5 mm

±1 mm

±2 mm

±3 mm

Static resolution *3

0.25 μm

Linearity *4

±0.45 μm

±0.9 μm

±2.0 μm

±3.0 μm

Spot diameter (Total measurent
 range) *5

50 μm dia.

70 μm dia.

100 μm dia.

120 μm dia.

Measurement cycle *6

20 μs to 400 μs

Operating ambient illumination

Illumination on object surface max.30000 Lx: (incandescent light)

Ambient temperature range

Operation: 0 to 50°C, Storage: -15 to +60°C
 (No freezing and condensation)

Ambient humidity range

Operation/storage: 35 or 85%RH (No condensation)

Degree of protection

IP40 (IEC60529)

Vibration resistance (destructive)

10 to 150 Hz (half amplitude 0.35 mm), 80 mins in each of X/Y/Z directions

Shock resistance (destructive)

150 m/s2, 6 direction, 3 times each (up/down, left/right, forward/backward)

Temperature characteristic *7

0.6 μm/°C
 (0.2 μm/°C)

1.1 μm/°C
 (0.5 μm/°C)

1.8 μm/°C
 (1.0 μm/°C)

2.1 μm/°C
 (1.2 μm/°C)

LED Safety

Risk Group 1 (IEC62471)

Material

Chassis: aluminum die cast
 Fiber cable sheath: PVC
 Calibration ROM: PC

Fiber cable length

0.3 m, 2 m (flex-resistant cable)

Fiber cable minimum bend radius

20 mm

Insulation resistance
 (Calibration ROM)

Between case and all terminals: 20 MΩ (by 250 VDC)

Dielectric strength
 (Calibration ROM)

Between case and all terminals: 1000 VAC, 50/60 Hz, 1 min

Weight

Fiber cable length 0.3 m Approx. 170g
 Fiber cable length 2 m Approx. 180g

Accessories

Calibration ROM fixing screw (M2×5mm) × 1,
 Fiber cable protective cap × 1, Strap × 2,
 Instruction Manual, Precautions

Item

Specifications

ZW-SP7007

ZW-SP7010

ZW-SPR7007

ZW-SPR7010

Sensor controller

ZW-7000T

Sensor head type

Pen-shaped straight type

Pen-shaped right angle type

Measurement center distance *1

7 mm

10 mm

7 mm

10 mm

Measuring range *2

±0.3 mm

±0.7 mm

±0.3 mm

±0.7 mm

Static resolution *3

0.25 μm

Linearity *4

±0.45 μm

±0.7 μm

±0.7 μm

±1.1 μm

Spot diameter (Total measurent
 range) *5

130 μm dia.

170 μm dia.

150 μm dia.

190 μm dia.

Measurement cycle *6

20 μs to 400 μs

Operating ambient illumination

Illumination on object surface max.30000 Lx: (incandescent light)

Ambient temperature range

Operation: 0 to 50°C, Storage: -15 to +60°C
 (No freezing and condensation)

Ambient humidity range

Operation/storage: 35 or 85%RH (No condensation)

Degree of protection

IP40 (IEC60529)

Vibration resistance (destructive)

10 to 150 Hz (half amplitude 0.35 mm), 80 mins in each of X/Y/Z directions

Shock resistance (destructive)

150 m/s2, 6 direction, 3 times each (up/down, left/right, forward/backward)

Temperature characteristic *7

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

LED Safety

Risk Group 1 (IEC62471)

Material

Chassis: SUS
 Fiber cable sheath: PVC
 Calibration ROM: PC
 Mounting Plate: Aluminum

Chassis: SUS, aluminum
 Fiber cable sheath: PVC
 Calibration ROM: PC
 Mounting Plate: Aluminum

Fiber cable length

0.3 m, 2 m (flex-resistant cable)

Fiber cable minimum bend radius

20 mm

Insulation resistance
 (Calibration ROM)

Between case and all terminals: 20 MΩ (by 250 VDC)

Dielectric strength
 (Calibration ROM)

Between case and all terminals: 1000 VAC, 50/60 Hz, 1 min

Weight

Fiber cable length 0.3 m Approx. 27 g
 Fiber cable length 2 m Approx. 37 g

Fiber cable length 0.3 m Approx. 31 g
 Fiber cable length 2 m Approx. 41 g

Accessories

Installation plate × 1, Unit fixing screws (M2 × 10 mm) × 4,
 Calibration ROM fixing screw (M2 × 5 mm) × 1,
 Fiber cable protective cap × 1, Strap × 2,
 Instruction Manual, Precautions

1. Indicates the distance from the front of the sensor head. The pen-shaped right angle type has a maximum individual
       difference of ±0.15 mm in the distance from the front of the sensor head.
 2. The measurement range is higher 28 μs than measurement cycle.
 3. Capacity value when OMRON standard mirror surface target is measured at the measurement center distance as the
       average of 16,384 times.
       The value when the Sensor Controller ZW-7000T is connected.
 4. Material setting for the OMRON standard mirror surface target: Error from an ideal straight line when measuring on
       mirror surface.
5. Capacity value defined by 1/e2 (13.5%) of the peak optical intensity of the measurement wavelength.
 6. When an extension fiber cable of 10 m or longer is connected, the setting rage of the measurement cycle (exposure
       time) changes. For details, refer to Setting Measurement Cycle in the ZW-8000/7000/5000 User's Manual (Cat. No.
       Z362).
 7. Actual value of the change in measurement value at the measurement center distance when fastened with an
      aluminum jig between the Sensor Head and the target, and with the Sensor Head and the Sensor Controller set in the
      same temperature environment.
      The value in parentheses is the actual value when using an SUS304 jig.
      When measuring the thickness, the value is calculated from the difference between the heights of the surface and rear
      surface, so there is no effect on the temperature change.

ZW-S5010/S5020/S5030/SP5007/SP5010/SPR5007/SPR5010

Item

Specifications

ZW-S5010

ZW-S5020

ZW-S5030

Sensor controller

ZW-5000T

Sensor head type

Square-shaped straight type

Measurement center distance *1

10 mm

20 mm

30 mm

Measuring range

±0.5 mm

±1 mm

±2 mm

Static resolution *2

0.25 μm

Linearity *3

±0.45 μm

±0.9 μm

±2.0 μm

Spot diameter (Total measurent
 range) *4

9 μm dia.

13 μm dia.

18 μm dia.

Measurement cycle *5

80 μs to 1,600 μs

Operating ambient illumination

Illumination on object surface max.30000 Lx: (incandescent light)

Ambient temperature range

Operation: 0 to 50°C, Storage: -15 to +60°C
 (No freezing and condensation)

Ambient humidity range

Operation/storage: 35 or 85%RH (No condensation)

Degree of protection

IP40 (IEC60529)

Vibration resistance (destructive)

10 to 150 Hz (half amplitude 0.35 mm), 80 mins in each of X/Y/Z directions

Shock resistance (destructive)

150 m/s2, 6 direction, 3 times each (up/down, left/right, forward/backward)

Temperature characteristic *6

0.6 μm/°C
 (0.2 μm/°C)

1.1 μm/°C
 (0.5 μm/°C)

1.8 μm/°C
 (1.0 μm/°C)

LED Safety

Risk Group 1 (IEC62471)

Material

Chassis: aluminum die cast
 Fiber cable sheath: PVC
 Calibration ROM: PC

Fiber cable length

0.3 m, 2 m (flex-resistant cable)

Fiber cable minimum bend radius

20 mm

Insulation resistance
 (Calibration ROM)

Between case and all terminals: 20 MΩ (by 250 VDC)

Dielectric strength
 (Calibration ROM)

Between case and all terminals: 1000 VAC, 50/60 Hz, 1 min

Weight

Fiber cable length 0.3m Approx. 170g
 Fiber cable length 2m Approx. 180g

Accessories

Calibration ROM fixing screw (M2×5mm) × 1,
 Fiber cable protective cap × 1, Strap × 1,
 Instruction Manual, Precautions

Item

Specifications

ZW-SP5007

ZW-SP5010

ZW-SPR5007

ZW-SPR5010

Sensor controller

ZW-5000T

Sensor head type

Pen-shaped straight type

Pen-shaped right angle type

Measurement center distance *1

7 mm

10 mm

7 mm

10 mm

Measuring range

±0.3 mm

±0.7 mm

±0.3 mm

±0.7 mm

Static resolution *2

0.25 μm

Linearity *3

±0.45 μm

±0.7 μm

±0.7 μm

±1.1 μm

Spot diameter (Total measurent
 range) *4

13 μm dia.

18 μm dia.

15 μm dia.

20 μm dia.

Measurement cycle *5

80 μs to 1,600 μs

Operating ambient illumination

Illumination on object surface max.30000 Lx: (incandescent light)

Ambient temperature range

Operation: 0 to 50°C, Storage: -15 to +60°C
 (No freezing and condensation)

Ambient humidity range

Operation/storage: 35 or 85%RH (No condensation)

Degree of protection

IP40 (IEC60529)

Vibration resistance (destructive)

10 to 150 Hz (half amplitude 0.35 mm), 80 mins in each of X/Y/Z directions

Shock resistance (destructive)

150 m/s2, 6 direction, 3 times each (up/down, left/right, forward/backward)

Temperature characteristic *6

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

0.8 μm/°C
 (0.4 μm/°C)

LED Safety

Risk Group 1 (IEC62471)

Material

Chassis: SUS
 Fiber cable sheath: PVC
 Calibration ROM: PC
 Mounting Plate: Aluminum

Chassis: SUS, aluminum
 Fiber cable sheath: PVC
 Calibration ROM: PC
 Mounting Plate: Aluminum

Fiber cable length

0.3 m, 2 m (flex-resistant cable)

Fiber cable minimum bend radius

20 mm

Insulation resistance
 (Calibration ROM)

Between case and all terminals: 20 MΩ (by 250 VDC)

Dielectric strength
 (Calibration ROM)

Between case and all terminals: 1000 VAC, 50/60 Hz, 1 min

Weight

Fiber cable length 0.3 m Approx. 29 g
 Fiber cable length 2 m Approx. 39 g

Fiber cable length 0.3 m Approx. 33g
 Fiber cable length 2 m Approx. 43g

Accessories

Installation plate × 1, Unit fixing screws (M2 × 10 mm) × 4,
 Calibration ROM fixing screw (M2 × 5 mm) × 1,
 Fiber cable protective cap × 1, Strap × 1,
 Instruction Manual, Precautions







1. Indicates the distance from the front of the sensor head. The pen-shaped right angle type has a maximum individual
       difference of ±0.15 mm in the distance from the front of the sensor head.
 2. Capacity value when OMRON standard mirror surface target is measured at the measurement center distance as the
       average of 16,384 times.
       The value when the Sensor Controller ZW-5000T is connected.
 3. Material setting for the OMRON standard mirror surface target: Error from an ideal straight line when measuring on
       mirror surface.
4. Capacity value defined by 1/e2 (13.5%) of the peak optical intensity of the measurement wavelength.
 5. When an extension fiber cable of 5 m or longer is connected, the setting rage of the measurement cycle (exposure
       time) changes. For details, refer to Setting Measurement Cycle in the ZW-8000/7000/5000 User's Manual (Cat. No.
       Z362).
 6. Actual value of the change in measurement value at the measurement center distance when fastened with an
      aluminum jig between the Sensor Head and the target, and with the Sensor Head and the Sensor Controller set in the
      same temperature environment.
      The value in parentheses is the actual value when using an SUS304 jig.
      When measuring the thickness, the value is calculated from the difference between the heights of the surface and rear
      surface, so there is no effect on the temperature change.

Sensor Controller

Item

Specifications

ZW-8000T

ZW-7000T

ZW-5000T

Input/output type

NPN/PNP dual type

Number of connected sensor heads

1

Sensor head compatibility

ZW-S80[][]/
 ZW-SP80[][]/
 ZW-SPR80[][]

ZW-S70[][]/
 ZW-SP70[][]/
 ZW-SPR70[][]

ZW-S50[][]/
 ZW-SP50[][]/
 ZW-SPR50[][]

LED Safety

Risk Group 1 (IEC62471)

LASER safety

Class1 (IEC/EN60825-1)

Segment
 Display

Main display

11-segment white display, 6 digits

Sub-display

11-segment green display, 6 digits

LED
 display

Status indicators

HIGH (orange), PASS (green), LOW (orange), STABILITY (green),
 ZERO (green), ENABLE (green), THRESHOLD-H (orange),
 THRESHOLD-L (orange), RUN (green)

EtherCAT indicator

ECAT RUN (green), L/A IN (Link/Activity IN) (green), L/A OUT
 (Link/Activity OUT) (green), ECAT ERR (red)

External
 I/F

Ethernet

100BASE-TX/10BASE-T, Non-procedure (TCP/UDP), EtherNet/IP

EtherCAT

EtherCAT exclusive protocol 100BASE-TX

RS-232C

Max. 115,200 bps

Analog
 output
 terminal
 block

Analog voltage output
 (OUT V)

-10 V to +10 V, output impedance: 100 Ω

Analog current output
 (OUT A)

4 mA to 20 mA, max. load resistance: 300 Ω

32-pole
 expansion
 connector

Judgment output
 (HIGH/PASS/LOW)

Transistor output system
 Output voltage: 21.6 to 30 VDC
 Load current: 50 mA or less
 Residual voltage when turning ON: 2 V or less
 Leakage voltage when turning OFF: 0.1 mA or less

Busy output (BUSY)

Alarm output (ALARM)

Enable output
 (ENABLE)

Sync flag output
 (SYNFLG)

Trigger busy output
 (TRIGBUSY)

Logging state output
 (LOGSTAT)

Logging error output
 (LOGERR)

Stability output
 (STABILITY)

Task state output
 (TASKSTAT)

LIGHT OFF input
 (LIGHT OFF)

DC input system
 Input voltage: 24 VDC ± 10% (21.6 to 26.4 VDC)
 Input current: 7 mA Type. (24 VDC)
 ON voltage/ON current: 19 V/3 mA or less
 ON voltage/ON current: 5 V/1 mA or less

Zero reset input
 (ZERO)

Timing input (TIMING)

Reset input (RESET)

Sync input (SYNC)

Trigger input (TRIG)

Logging input
 (LOGGING)

Bank

Currently
 selected
 bank output
 (BANK_OUT
 1 to 3)

Transistor output system
 Output voltage: 21.6 to 30 VDC
 Load current: 50 mA or less
 Residual voltage when turning ON: 2 V or less
 Leakage voltage when turning OFF: 0.1 mA or less

Bank Selection
 input
 (BANK_SEL
 1 to 3)

DC input system
 Input voltage: 24 VDC ± 10% (21.6 to 26.4 VDC)
 Input current: 7 mA Type. (24 VDC)
 ON voltage/ON current: 19 V/3 mA or more
 OFF voltage/OFF current: 5 V/1 mA or less

Main
 functions

Exposure time

Automatic/Fixed

Measuring cycle *1

60 μs to 7,500 μs

20 μs to 400 μs

80 μs to 1,600 μs

Material setting

Standard/Mirror/Rough surfaces

Measurement item

Height/Thickness of transparent object/Calculation

Filtering

Median/Average/Differentiation/High pass/Low pass/Band pass

Output

Scaling/Different holds/Zero reset/Logging for a measured value/
 Keep, Clamp

Display

Measured value/Threshold value/Analog output voltage or current
 value/Judgment result/Resolution/Light power/Internal logging
 condition/Peak amount of received light

Number of configurable banks

NORMAL mode: Max. 8 banks
 JUDGMENT mode: Max. 32 banks

Task process

Multi-task (up to 4 tasks per bank)

System

Save/Initialization/Display measured information/Communication
 settings/Sensor head calibration/Key-lock/Zero reset memory/
 Timing input

Rating

Power supply voltage

21.6 to 26.4 VDC (including ripple)

Current consumption

700 mA or less

800 mA or less

Insulation resistance

Across all lead wires and FG terminal: 20 MΩ (by 250 VDC)

Dielectric strength

Between all lead wires and FG terminal: 500 VAC, 50/60 Hz,
 1 minute

Environ-
 mental
 resistance

Degree of protection

IP20 (IEC60529)

Vibration resistance (destructive)

10 to 55 Hz (half amplitude 0.35 mm), 50 mins in each of X/Y/Z
 directions

Shock resistance (destructive)

150 m/s2, 6 direction, 3 times each (up/down, left/right, forward/
 backward)

Ambient temperature range

Operation: 0 to 40°C,
 Storage: -15 to +60°C (No freezing and condensation)

Ambient humidity range

Operation/storage: 35 to 85%RH (No condensation)

Grounding

D-type grounding (grounding resistance of 100 Ω or less)
 Note: For conventional Class D grounding

Material

Chassis: PC

Weight

Approx. 950 g
 (main unit only),
 Approx. 150 g
 (Parallel cable)

Approx. 900g (main unit only),
 Approx. 150 g (Parallel cable)

Accessories

Parallel cable (ZW-XCP2E) × 1
 10 Fiber cleaners (ZW-XCL) × 1
 Instruction Manual
 Member registration sheet
 Precautions

Parallel cable
 (ZW-XCP2E) × 1
 10 Fiber cleaners
 (ZW-XCL) × 1
 Fiber adapter cap × 1
 Strap × 1
 Instruction Manual
 Member registration
 sheet
 Precautions

Note: The Export Trade Control Order compatible Sensor Controller (ZW-8000T/7000T/5000T) is available.
          When using this Controller, the minimum resolution is 0.25 μm regardless of the connected Sensor Head and setting
          conditions.
 1. When an extension fiber cable of 2 m or longer (on the ZW-8000 series), 10 m or longer (on the ZW-7000 series) or
      5 m or longer (on the ZW-5000 series) is connected, the setting rage of the measurement cycle (exposure time)
      changes. For details, refer to Setting Measurement Cycle in the ZW-8000/7000/5000 User's Manual (Cat. No. Z362).

EtherCAT Communications Specifications

Item

Specification

Communications standard

IEC61158 Type12

Physical layer

100BASE-TX(IEEE802.3)

Connectors

RJ45 × 2
 ECAT IN: EtherCAT input
 ECAT OUT: EtherCAT output

Communications media

Category 5 or higher (cable with double, aluminum tape and braided shielding) is
 recommended.

Communications distance

Distance between nodes: 100 m max.

Process data

Variable PDO mapping

Mailbox (CoE)

Emergency messages, SDO requests, SDO responses, and SDO information

Distributed clock

Synchronization in DC mode.

LED display

L/A IN (Link/Activity IN) × 1, AL/A OUT (Link/Activity OUT) × 1, AECAT RUN × 1,
 AECAT ERR × 1

Automation Software Sysmac Studio

Item

Operating environment *3

Operating system (OS)
 *1

Windows 7 SP1 (32-bit/64-bit version)/Windows 8.1 (32-bit/64-bit version)/
 Windows 10(32-bit/64-bit version)/Windows 11 (64-bit version)

CPU

Windows computers with Intel® Celeron® processor 540 (1.8 GHz) or faster CPU.
 Intel® Core™ i5 M520 processor (2.4 GHz) or equivalent or faster recommended.

Main memory

2 GB min.
 4 GB min. recommended

Hard disk

Minimum 4.6 GB of Hard disk space is required to install. *2

Display

XGA 1024 × 768, 16,000,000 colors
 WXGA 1280 × 800 dots or higher resolution is recommended.

Disk drive

DVD-ROM drive

Communications ports

USB port corresponded to USB 2.0, or Ethernet port *4

Supported languages

Japanese, English, German, French, Italian, Spanish, simplified Chinese,
 traditional Chinese, Korean

1. Note about Sysmac Studio compatible operating systems: The required system and hard disk capacity differs according
       to the system environment.
 2. Separate logging memory is required to use the file logging function.
 3. Describes System Requirements and notes of Sysmac Studio Measurement Sensor Edition.
       For details on System Requirements and notes of Sysmac Studio Measurement Sensor Edition, refer to Sysmac Studio
       Version 1 Operation Manual.
 4. For information on how to connect a personal computer with the controller or other hardware and information on
      required cables, refer to manuals for each hardware.

Version Information

Sensor Head/Cable, Sensor Controller, and Sysmac Studio

The applicable version of the Sensor Controller varies depending on the Sensor Head or Cable. The versions are listed below.
 Use the latest version of Sysmac Studio Standard Edition/Measurement Sensor Edition. 

Sensor head/Cable

ZW Series

Version of Sensor
 Controller

Corresponding version of
 Sysmac Studio Standard
 Edition/Measurement
 Sensor Edition

Type

Model

Square-shaped straight type

ZW-S80[]0 []M

ZW-8000[]

Version 3.000 or later

Version 1.22 or higher

Pen-shaped straight type

ZW-SP8007 []M
 ZW-SP8010 []M

Pen-shaped right-angle type

ZW-SPR8007 []M
 ZW-SPR8010 []M

Extension Fiber Cable

ZW-XF80[][]R

Square-shaped straight type

ZW-S70[]0 []M

ZW-7000[]

Version 2.030 or later

Version 1.15 or higher

Pen-shaped straight type

ZW-SP7007 []M
 ZW-SP7010 []M

Version 2.110 or later

Pen-shaped right-angle type

ZW-SPR7007 []M
 ZW-SPR7010 []M

Extension Fiber Cable

ZW-XF7002R
 ZW-XF7005R

Version 2.030 or later

ZW-XF7010R
 ZW-XF7020R
 ZW-XF7030R

Version 2.100 or later

Square-shaped straight type

ZW-S50[]0 []M

ZW-5000[]

Version 2.100 or later

Version 1.18 or higher

Pen-shaped straight type

ZW-SP5007 []M
 ZW-SP5010 []M

Version 2.110 or later

Pen-shaped right-angle type

ZW-SPR5007 []M
 ZW-SPR5010 []M

Extension Fiber Cable

ZW-XF50[][]R

Version 2.100 or later

Note: Refer to the Firmware Update in the ZW-8000/7000/5000 User's Manual (Cat. No. Z362) for how to update the
          Sensor Controller.