Smart Sensors Laser Displacement Sensors CMOS Type

ZX2

Stable measurement that is unaffected by workpiece changes. The simple setting for everyone.

Diffuse-reflective Sensor Heads

Model

ZX2-LD50L

ZX2-LD50

ZX2-LD100L

ZX2-LD100

Optical system

Diffuse reflective

Light source
 (wave length)

Visible-light semiconductor laser with a wavelength of 660 nm and an output of 1mW max.

EN class 2, FDA class 1 *5

Measurement center
 point

50 mm

100 mm

Measurement range

±10 mm

±35 mm

Beam shape

Line

Spot

Line

Spot

Beam size *1

Approx. 60 μm ×
 2.6 mm

Approx. 60 μm dia.

Approx.110 μm ×
 2.7 mm

Approx.110 μm dia.

Resolution *2

1.5μm

5μm

Linearity *3

±0.05% F.S.
 (40 to 50 mm)

±0.1% F.S.
 (40 to 50 mm)

±0.05% F.S.
 (65 to 100 mm)

±0.1% F.S.
 (65 to 100 mm)

Temperature
 characteristic *4

0.02% F.S./°C

Ambient illumination

Incandescent lamp: 10,000 lx max. (on light receiving side)

Ambient temperature

Operating: 0 to 50 °C, Storage: -15 to 70 °C (with no icing or condensation)

Ambient humidity

Operating and storage: 35% to 85% (with no condensation)

Dielectric strength

1,000 VAC, 50/60 Hz for 1 min.

Vibration resistance
 (destruction)

10 to 150 Hz, 0.7-mm double amplitude, 80 min. each in X, Y, and Z directions

Shock resistance
 (destruction)

300 m/s2 3 times each in six directions (up/down, left/right, forward/backward)

Degree of protection

IEC60529, IP67

Connection method

Connector connection (standard cable length: 500 mm)

Weight (packed state)

Approx. 160 g (main unit only: Approx. 75 g)

Materials

Case and cover: PBT (polybutylene terephtahalate), Optical window: Glass, Internal thread: Brass, Cable: PVC

Accessories

Instruction sheet, Ferrite core ×1 (made by TDK Corp. ZCAT1730-0730A), Laser warning
 label (English), FDA

Regular-reflective Sensor Head

Model

ZX2-LD50L

Optical system

Regular reflective

Light source (wave length)

Visible-light semiconductor laser with a wavelength of 660 nm and an output of 0.24
 mW max. EN class 1, FDA class 1 *5

Measurement center point

48 mm

Measurement range

±5 mm

Beam shape

Spot

Beam size *1

Approx. 60 μm dia.

Resolution *2

1.5 μm

Linearity *3

±0.3%F.S. (entire range)

Temperature
 characteristic *4

0.06%F.S./°C

Ambient illumination

Incandescent lamp: 10,000lx max. (on light receiving side)

Ambient temperature

Operating: 0 to 50 °C, Storage: -15 to 70 °C (with no icing or condensation)

Ambient humidity

Operating and storage: 35% to 85% (with no condensation)

Dielectric strength

1,000 VAC, 50/60 Hz for 1 minute

Vibration resistance
 (destruction)

10 to 150 Hz, 0.7-mm double amplitude, 80 minutes, each in X,Y,and Z directions

Shock resistance
 (destruction)

300 m/s2 3 times each in six directions (up/down,left/right,forward/backward)

Degree of protection

IEC 60529, IP67

Connection method

Connector connection (standard cable length: 500 mm)

Weight (packed state)

Approx.160g (Sensor Head only: Approx.75g)

Materials

Case and cover: PBT (polybutylene terephtahalate), Optical window: Glass, Internal thread: Brass, Cable: PVC

Accessories

Instruction sheet, Ferrite core ×1 (made by TDK Corp. ZCAT1730-0730A), Laser
 warning label (English), FDA

Note: False detection outside the measurement range can occur in the case of an object with high reflectance.
1. Beam size: Defined as 1/e2 (13.5 %) of the center optical intensity at the minimum value of the measurement range (effective value).
 False detections can occur in the case there is light leakage outside the defined region and the surroundings of the target object have a high reflectance in comparison to the target object. Correct measurements may not be obtained if the workpiece is smaller than the beam size.
 2. Resolution: indicates the degree of fluctuation (±3σ) of analog output when connected to the ZX2-LDA. (The measured value is given for the center distance for OMRON’s standard target object (diffuse-reflective models: white ceramic object, regular-reflective models:1/4 λ flat mirror) when the response time of the ZX2-LDA is set to 128 ms.) Indicates the repetition accuracy for when the workpiece is in a state of rest. Not an indication of distance accuracy. Resolution performance may not be satisfied in a strong electromagnetic field.
 3. Linearity: indicates the error with respect to the ideal straight line of the displacement output in the case of measuring Omron's standard target object. Linearity and measured value may vary depending on target object. F.S. indicates the full scope of the measurement range. (ZX2-LD50(L): 20 mm)
 4. Temperature characteristic: Value for the case the space between the sensor head and Omron’ s standard target object is secured by an aluminum jig. (Measured at the measurement center distance)
 5. These Sensors are classified as Class 1 under EN 60825-1 and the regulations of Laser Notice No. 56 for FDA certification. CDRH registration has been completed.

Amplifier Units

Model

ZX2-LDA11

ZX2-LDA41

Measurement period *1

Min. 30 μs

Response time

60 μs, 120 μs, 240 μs, 500 μs, 1 ms, 2 ms, 4 ms, 8 ms, 12 ms, 20 ms, 36 ms, 66 ms,
 128 ms, 250 ms, 500ms

Analog output *2

4 to 20 mA, Max. load resistance: 300Ω/±5 VDC or 1 to 5 VDC, Output impedance: 100Ω

Judgement outputs
 (HIGH/PASS/LOW: 3
 outputs), error output

NPN open-collector outputs, 30 VDC, 50
 mA max. (residual voltage: 1 V max. for load
 current 10 mA max., 2V max. for load
 current above 10 mA)

PNP open-collector outputs, 30 VDC, 50
 mA max. (residual voltage: 1 V max. for load
 current 10 mA max., 2 V max. for load
 current above 10 mA)

Laser OFF input, zero
 reset input, timing input,
 reset input, bank input

ON: Short-circuited with 0-V terminal or
 1.2 V or less
 OFF: Open (leakage current: 0.1 mA max.)

ON: Supply voltage short-circuited or supply
 voltage within -1.2 V
 OFF: Open (leakage current: 0.1 mA max.)

Functions

Smart tuning, scaling, sample hold, peak hold, bottom hold, peak-to-peak hold, self-peak
 hold, self-bottom hold, average hold, zero reset, On-delay timer, OFF-delay timer, keep/
 clamp switch, (A-B) calculations *3, thickness calculation *3, mutual interference
 prevention *3, laser deterioration detection, bank function (4 banks), differential function

Indications

Judgement indicators: HIGH (orange), PASS (green), LOW (orange), 11-segment main
 display (red), 11-segment sub-display (orange), laser ON (green), zero reset (green),
 enable (green), menu (green), HIGH threshold (orange), LOW threshold (orange)

Power supply voltage

10 to 30 VDC, including 10% ripple (p-p)

Power consumption

3,000 mW max. (at 24 VDC: 125 mA max., at 12 VDC: 250 mA max.)

Ambient temperature

Operating: 0 to 50 °C, Storage: -15 to 70 °C (with no icing or condensation)

Ambient humidity

Operating and storage: 35% to 85% (with no condensation)

Dielectric strength

1,000 VAC, 50/60 Hz for 1 min.

Vibration resistance
 (destruction)

10 to 150 Hz, 0.7-mm double amplitude, 80 min. each in X, Y, and Z directions

Shock resistance
 (destruction)

300 m/s2 3 times each in six directions (up/down, left/right, forward/backward)

Degree of protection

IEC60529, IP40

Connection method

Prewired (standard cable length: 2 m)

Weight (packed state)

Approx. 200 g (main unit only: Approx. 135 g)

Materials

Case: PBT (polybutylene terephtahalate), Cover: Polycarbonate, Display: Methacrylic resin,
 Button: Polyacetal, Cable: PVC

Accessories

Instruction sheet

1. In the case of Omron’s standard target object (white ceramic)
 2. Select current output (4 to 20 mA) and voltage output (±5V or 1 to 5V) by MENU mode.
 3. Calculating unit (ZX2-CAL) is necessary. Calculations are possible for up to two amplifier units. Mutual interference prevention is possible for up to five amplifier units.

Calculating Unit

Model

ZX2-CAL

Applicable Amplifier Units

ZX2-LDA11, ZX2-LDA41

Current consumption

12 mA max. (supplied from the Smart Sensor Amplifier Unit)

Ambient temperature

Operating: 0 to +50°C, storage: -15 to +70°C (with no icing or condensation)

Ambient humidity

Operating and storage: 35% to 85% (with no condensation)

Connection method

Connector

Dielectric strength

1,000 VAC, 50/60 Hz for 1 minute

Vibration resistance (destruction)

10 to 150 Hz, 0.7-mm double amplitude, 80 minutes, each in X,Y,and Z directions

Shock resistance (destruction)

300 m/s2 3 times each in six directions (up/down, left/right, forward/backward)

Materials

Case: ABS, Display: Methacrylic resin

Weight (packed state)

Approx. 50g (Calculating Unit only: Approx. 15g)

Accessories

Instruction sheet

ZX2-series Communications Interface Unit

Model

ZX2-SF11

Power supply voltage

10 to 30 V DC ±10% (including 10% ripple (p-p))
 (Supplied from Sensor Amplifier.)

Power consumption

720 mW max. (at 24 V: 30 mA max., at 12 V: 60 mA max.)
 (Not including Sensor Amplifier current consumption or output current

Applicable Amplifier Units

ZX2-LDA[][] (Production after November 2013)

Applicable Amplifier
 Unit versions

Sensor Amplifier Unit version: V1.330 or higher
 (The Sensor Amplifier version is shown on the sub-digital
 display when the power of the Sensor Amplifier is turned ON.)

Max. No. of Amplifier
 Units

5

Commu-
 nications
 functions

Port

RS-232C (9-pin, D-Sub connector)

Communications method

Full duplex

Synchronization method

Start/stop synchronization

Transmission code

ASCII

Baud rate

38,400 (at shipping)/9,600 bps switchable

Data bit length

8 bits

Parity check

None

Stop bit length

1 bit

Data
 delimiter

Receiving

CR or CR + LF is automatically recognized.

Sending

CR + LF fixed

Indicators

Power supply: green, Sensor communications: green, Sensor
 communications error: red, External terminal communications:
 green, External terminal communications error: red

Protective circuits

Power supply reverse polarity protection

Ambient temperature

Operating: 0 to 50°C, storage: -15 to 60°C
 (with no icing or condensation)

Ambient humidity

Operating and storage: 35% to 85%
 (with no condensation)

Insulation resistance

20 MΩ min. (at 500 VDC)

Dielectric strength

1,000 VAC, 50/60 Hz for 1 min,
 Leakage current: 10 mA max.

Materials

Case: PBT (polybutylene terephthalate), Cover: Polycarbonate

Accessories

Instruction sheet, 2 clamps