Aluminum analyzer Liquiline System CA80AL

Colorimetric system for drinking water and wastewater monitoring and optimization of process control

Predecessor: Stamolys CA71AL

  • Measurement range: 15 to 1000 µg/l Al, 15 to 1000 µg/l with dilution function to maximum 300 to 20 000 µg/l Al

  • Process temperature: 4 … 40 °C (39.2 … 104 °F)

  • Process pressure: at atmospheric pressure, < 0.2 bar

Iron analyzer Liquiline System CA80FE

Colorimetric system for monitoring of drinking water, mineral water and process water

Predecessor: Stamolys CA71FE

  • Measurement range: 0.05 to 2 mg/l Fe
    0.1 to 5 mg/l Fe
    0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe

  • Process temperature: 4 to 40 °C (39.2 to 104 °F)

  • Process pressure: At atmospheric pressure, < 0.2 bar

Nitrite analyzer Liquiline System CA80NO

Colorimetric system for monitoring of drinking water, mineral water or raw water for food production

Predecessor: Stamolys CA71NO

  • Measurement range: 10 to 500 µg/l NO2-N
    0.2 to 3.0 mg/l NO2-N
    0.1 to 1.0 mg/l NO2-N
    0.1 to 1.0 mg/l with dilution function to maximum 5 to 50 mg/l NO2-N

  • Process temperature: 4 to 40 °C (39 to 104 °F)

  • Process pressure: At atmospheric pressure, < 0.2 bar absolute

TP analyzer Liquiline System CA80TP

Total phosphorus analyzer for environmental monitoring, industrial and municipal wastewater

Predecessor: SPECTRON TP CA72TP

  • Measurement range: 0 to 2 mg/l
    0.05 to 10 mg/l
    0.5 to 50 mg/l

  • Process temperature: 4 to 40 °C (39.2 to 104 °F)

  • Process pressure: At atmospheric pressure, < 0.2 bar

Flow assembly for chlorine dioxide sensors Flowfit CCA151

Simple assembly for drinking and process water applications

Successor: Flowfit CYA27

  • Process temperature: Max. 60 °C (Max. 140 °F)

  • Process pressure: Max. 4 bar relative(Max. 58 psi relative)

Analog free chlorine sensor CCS51

Sensor for drinking water, pool and process water as well as utilities in all industries

Predecessor: Chlorine CCS140, Trace Chlorine CCS141

  • Measurement range: Trace: 0 to 5 mg/l HOCl, Standard: 0 to 20 mg/l HOCl

  • Process temperature: +0 to 55 °C (32 to 130°F), non-freezing

  • Process pressure: 1 bar (14.5 psi)

  • Measuring method: – Closed, membrane covered measuring cell
    – Reduction of free chlorine at the cathode

Digital ozone sensor Memosens CCS58D

Memosens sensor for drinking water, wastewater and process water in all industries

  • Measurement range: 0 to 2 mg/l

  • Process temperature: +0 to 45 °C (32 to 110 °F), non freezing

  • Process pressure: 1 bar (14.5 psi), 2 bar (29 psi) absolute

  • Measuring method: – closed 2 electrode system
    – conversion of ozone into a signal current in nA
    – signal current is proportional to the concentration of ozone
    – the measurement is not pH dependent

Conductivity compact device Smartec CLD134

Toroidal conductivity measuring system for hygienic and sterile applications

Predecessor: CLD130, CLS51

  • Measurement range: 100 µS/cm to 2000 mS/cm, Cell constant k : 6.3 1/cm

  • Process temperature: Max. 125 °C (max. 257 °F)
    For sterilization: up to 150 °C for max. 60 min (up to 302 °F for max. 60 min)

  • Process pressure: 12 bar abs up to 90 °C
    (174 psi abs up to 194 °F)

Digital conductivity sensor Indumax CLS54D

Memosens toroidal conductivity sensor for hygienic and sterile applications

  • Measurement range: 100 µS/cm to 2000 mS/cm, Cell constant k: 6.3 1/cm

  • Process temperature: -10 °C to 125 °C (14 °F to 257 °F)
    Sterilization: 150 °C at 6 bar abs for max. 60 min (302 °F at 87 psi abs for max. 60 min)

  • Process pressure: 13 bar abs up to 90 °C (188.5 psi abs up to 194 °F)
    9 bar at 125 °C (130.5 psi abs at 257 °F)

Digital oxygen sensor Oxymax COS22D

Memosens amperometric oxygen sensor for standard, hygienic and sterile applications

Predecessor: Oxymax COS21D, Oxymax W COS71

Successor: Memosens COS22E

  • Measurement range: 0.001 to 20 mg/l, 0 to 200 %SAT, 0 to 400 hPa

  • Process temperature: -5 to 135 °C (23 to 275 °F)

  • Process pressure: 1 to 12 bar abs (14.5 to 174 psi)

Digital oxygen sensor Oxymax COS51D

Memosens amperometric oxygen sensor for water, wastewater and utilities

Predecessor: Oxymax W COS31

Successor: Memosens COS51E

  • Measurement range: 0.01 to 100 mg/l, 0.00 to 1000 %SAT, 0 to 2000 hPa

  • Process temperature: -5 to 50 °C (20 to 120 °F)

  • Process pressure: max. 10 bar (max. 145 psi)

Zero-point gel COY8

Ready-to-use gel for validation, calibration, and adjustment of oxygen and disinfection sensors

Digital pH sensor Ceragel CPS71D

Memosens glass electrode for the chemical and life sciences industries

Successor: Memosens CPS61E, Memosens CPS71E

  • Measurement range: pH 0 to 14

  • Process temperature: 0 to 140 °C (32 to 284 °F)

  • Process pressure: 1 to 14 bar (15 to 203 psi)

Suspended solids sensor Turbimax CUS51D

Memosens sensor for suspended solids and turbidity measurement in water, wastewater and utilities

Predecessor: Turbimax W CUS41, Turbimax CUS65

  • Measurement range: For formazin: 0 to 4000 FNU, Display range up to 9999 FNU, For sludge: 0 to 300 g/L, The achievable measuring range depends very much on the media

  • Process temperature: Process temperature: -5 to 50 °C (23 to 122 °F)
    Ambient temperature: -20 to 60 °C (-4 to 140 °F)

Immersion assembly Flexdip CYA112

Modular assembly for water, wastewater and utilities

Predecessor: CPA530, CPA610, Dipfit W CYA611

  • Process temperature: 0 to 60 °C (32 to 140 °F)

Modular holder system Flexdip CYH112

Holder for in-situ measurement in water, wastewater and utilities

Predecessor: Universal CYH101

  • Process temperature: -20 to 60 °C (-4 to 140 °F)

Digital pH sensor Orbipac CPF81D

Memosens compact electrode for wastewater, primaries and metal industries

Successor: Memosens CPF81E

  • Measurement range: pH 0 to 14

  • Process temperature: Version LH: 0 to 110 °C (32 to 230 °F)
    Version NN: 0 to 80 °C (32 to 170 °F)

  • Process pressure: 1 to 10 bar abs at 80 °C (15 to 145 psi at 176 °F)

Digital ORP sensor Orbipac CPF82D

Memosens compact electrode for wastewater, primaries and metal

Successor: Memosens CPF82E

  • Measurement range: -1500mV to +1500mV

  • Process temperature: 0 to 80 °C (32 to 170 °F)

  • Process pressure: 1 to 10 bar abs at 80 °C (15 to 145 psi abs at 176 °F)

Digital pH sensor Orbisint CPS11D

Memosens glass electrode for standard applications

Successor: Memosens CPS11E

  • Measurement range: pH 0 to 14

  • Process temperature: 0 to 135 °C (32 to 275 °F)

  • Process pressure: 1 to 17 bar (15 to 246 psi)

Digital ORP sensor Orbisint CPS12D

Memosens glass electrode for standard applications

Successor: Memosens CPS12E

  • Measurement range: -1500 mV to +1500 mV

  • Process temperature: -15 to 135 °C (5 to 275 °F)

  • Process pressure: 0 to 16 bar (0 to 232 psi)

Digital combined pH/ORP sensor Memosens CPS16D

Memosens glass electrode for standard applications

Successor: Memosens CPS16E

  • Process temperature: 0 to 135 °C (32 to 275 °F)

  • Process pressure: 1 to 17 bar (15 to 246 psi)

Digital pH sensor Memosens CPS31D

Memosens glass electrode for application in drinking water and swimming pools

Successor: Memosens CPS31E

  • Measurement range: pH 1 to 12

  • Process temperature: 0 to 80 °C (32 to 176 °F)

  • Process pressure: 1 to 4 bar (15 to 60 psi)

Digital non-glass pH sensor Ceramax CPS341D

Memosens enamel electrode for food and life sciences industries

  • Measurement range: pH -2 to 14

  • Process temperature: 0 to 140 °C (32 to 280 °F)

  • Process pressure: 0 to 7 bar abs (0 to 101.5 psi)

Digital pH sensor Ceraliquid CPS41D

Memosens glass electrode for the chemical and life sciences industries

Successor: Memosens CPS41E

  • Measurement range; pH 0 to 14

  • Process temperature: 0 to 135 °C (32 to 275 °F)

  • Process pressure: 1 to 11 bar abs (15 to 160 psi)

Digital ORP sensor Ceraliquid CPS42D

Memosens glass electrode for applications with fast-changing medium compositions or low conductivity

Successor: Memosens CPS42E

  • Measurement range: -1500 mV – +1500 mV

  • Process temperature: -15°C – 130°C

  • Process pressure: max. 8bar

Digital ORP sensor Ceragel CPS72D

Memosens glass electrode for the chemical and life sciences industries

Successor: Memosens CPS72E

  • Measurement range: -1500mV – +1500mV

  • Process temperature: -15°C – 140°C

  • Process pressure: max. 16bar

Digital combined pH/ORP sensor Memosens CPS76D

Memosens glass electrode for the chemical and life sciences industries

Successor: Memosens CPS76E

  • Measurement rangeL: pH 0-14

  • Process temperature: -15°C … 140°C (5°F – 284°F)

  • Process pressure: max. 13 bar (188 psi)depends on selected version

Digital pH sensor Orbipore CPS91D

Memosens glass sensor for heavily polluted media in chemical processes, paper or paint production

Successor: Memosens CPS91E

  • Measurement range: 0 to 14 pH

  • Process temperature: 0 to 110 °C (32 to 230 °F)

  • Process pressure: 1 to 14 bar (15 to 203 psi)

Digital ORP sensor Orbipore CPS92D

Memosens glass sensor for heavily polluted media in chemical processes, paper or paint production

Successor: Memosens CPS92E

  • Measurement range: -1500 mV – +1500 mV

  • Process temperature: max. 110°C (230°F)

  • Process pressure: max. 13bar (189psi)

Digital combined pH/ORP sensor Memosens CPS96D

Memosens glass sensor for heavily polluted media in chemical processes, paper or paint production

Successor: Memosens CPS96E

  • Measurement range: pH 0-14

  • Process temperature: max. 110°C (230°F)

  • Process pressure: max. 10 bar (145 psi)